Sterilization Pass Through Chamber

by PBSC
The HPV Pass Thru Chamber is an integrated through the wall transfer device for material transfers between different classification rooms where either an air particulate clean up or material surface bio-sterilization is needed before transfer. The HPV requires a Bioquell vapour generator (typically the Clarus L), to be attached either room side or in a separate technical area. The bio-decontamination chamber can be fully interfaced to room construction with closing fascia panels. The sterilization transfer chamber is delivered completely assembled, pre-wired and tested The automated process monitors all critical control points of the disinfection cycle. The high level disinfection cycle takes between 40-45 minutes (load dependent). The load will be decontaminated before transfer via a validated 6 log reduction vaporised sporicidal gassing disinfection cycle. The developed cycle is qualified with biological indicator challenges of Geobacillus stearothermphilus. The disinfection cycle is fully automatic with digital signal exchanges between the chamber PLC and vapour generator PLC to control the automatic configuration changes needed as the cycles phases progress.

Specification Details for Sterilization Pass Through Chamber (Hygienic Doors)

Please note that these product properties are not a replacement for the manufacturer's literature and it is always recommended that PBSC is consulted before specifying.